Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2016 - (Recall #: Z-1670-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

ADVIA Chemistry Triglyceride_2, concentrated; TRIG_c; Catalog # SMN # 10697575. For in vitro diagnostic use in the quantitative measurement of triglycerides in human serum and plasma on ADVIA Chemistry systems.

Product Classification:

Class II

Date Initiated: March 7, 2016
Date Posted: May 25, 2016
Recall Number: Z-1670-2016
Event ID: 73635
Reason for Recall:

Siemens Healthcare Diagnostics has confirmed the ADVIA Chemistry Triglyceride concentrated reagent (TRIG_c) does not meet Instructions For Use (IFU) linearity claims at the upper limit of the assay range as it approaches end of shelf life for lots # 348297 and 359932.

Status: Terminated
Product Quantity: 7,433 units (4,228 for lot 348297 and 3,199 for lot # 359932)
Code Information:

Kit lots 348297 & 359932.

Distribution Pattern:

Worldwide Distribution -- USA, including the states of CA, TN, UT, OK, FL, MI, MN, NY, CT, NC, KS, TX, CO, WA, NM, and OR; and, the countries of Austria, Australia, Belgium, Brazil, Bulgaria, Denmark, Estonia, France, Germany. Great Britain, Italy, Latvia, Netherland, Norway, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Czech Republic, India, New Zealand, Thailand, Turkey, and Russian Fed.

Voluntary or Mandated:

Voluntary: Firm initiated