Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2017 - (Recall #: Z-1120-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LK3PS1, SMN 10706293 (US Kit Catalog) Product Usage: For in vitro diagnostic use with the IMMULITE¿ and IMMULITE 1000 Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients.
Class II
Negative bias results with the IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit was confirmed for two lots (D0119 and D0120). The investigation by Siemens has confirmed that patient samples exhibit an average negative bias of approximately -22% with a range of -16% to -27% for samples recovering from 0.08 to 16.2 ng/mL.
Lot Numbers: 1) D01196 - Expiration date 2016-12-31 2) D0120 - Expiration date 2017-03-31
Worldwide Distribution - US Nationwide and the countries of Denmark, Brazil, Croatia and Germany
Voluntary: Firm initiated