Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2017 - (Recall #: Z-2166-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
Class II
Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8¿C, as published in the Instructions for Use (IFU). Adjustors stored at 2-8C after 14 days may show a bias in patient samples and quality control falling outside of published ranges. Lookback not recommended.
Lot Numbers: 1) 357, 2) 358, 3) 360, 4) 361 & 5) 363 Expiration Dates: 1) 2017-01-30, 2) 2017-03-31, 3) 2017-04-30, 4) 2017-05-31 & 5) 2017-06-30
Nationwide Distribution
Voluntary: Firm initiated