Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2017 - (Recall #: Z-2608-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Dimension Vista Estradiol (E2) Flex reagent cartridge
Class II
The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.
Device Listing Number # D121541 Lot # 16148AA, 16190BA, 16201BA, 16228BC, 16256BA, 16277BB 16298AB, 16343BE,17005BC
Product was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland India,Italy,Japan,Canary Islands,Republic of Korea,Kuwait,Lebanon,Latvia Macedonia,Netherlands,Norway,New Zealand,Philippines,Poland,Portugal Qatar,Saudi Arabia,Slovenia,Slovakia,Turkey,Taiwan,South Africa
Voluntary: Firm initiated