Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2017 - (Recall #: Z-2609-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Dimension Vista Ferritin (FERR) Flex reagent cartridge
Class II
The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.
Device Listing Number # D011479 Lot # 16102BF, 16144BB,16196BD,16201BF,16214BF,16236BA 16252AB,16277BC,16284AB,17012BC,17012BD,17034BA,16112BB 16148BA,16165BE,16195BA,16223AD,16259BC,16285BE,16312BD 16334BE,17004BA,17038BC
Product was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland India,Italy,Japan,Canary Islands,Republic of Korea,Kuwait,Lebanon,Latvia Macedonia,Netherlands,Norway,New Zealand,Philippines,Poland,Portugal Qatar,Saudi Arabia,Slovenia,Slovakia,Turkey,Taiwan,South Africa
Voluntary: Firm initiated