Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2018 - (Recall #: Z-1675-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi Anti HBc
Class II
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
Anti HBc, Anti HBc (IMMULITE 2000/IMMULITE 2000 XPi L2KHC2 (D)) Lot codes: Anti HBcM # 579 580 581 D579 D580 D581 ***Added 9/11/20*** D602
Nationwide and Canada, Mexico, and Thailand
Voluntary: Firm initiated