Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2018 - (Recall #: Z-1687-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
ADVIA Centaur¿ Cyclosporine
Product Classification:
Class II
Date Initiated: January 2, 2018
Date Posted: May 16, 2018
Recall Number: Z-1687-2018
Event ID: 79505
Reason for Recall:
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere
Status: Terminated
Product Quantity: N/A
Code Information:
ADVIA Centaur¿ Cyclosporine; Cyclosporine; SMN 10335448 Lot codes: ADVIA Centaur CSA Lot # 38182025 64432025 98554025 58299030 25547030 30153031 77934031 97170031
Distribution Pattern:
Nationwide and Canada, Mexico, and Thailand
Voluntary or Mandated:
Voluntary: Firm initiated