Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2018 - (Recall #: Z-2182-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
Dimension Vista Gentamicin Flex reagent cartridge, DV GENT, gentamicin enzyme immunoassay. Affected lots are 17135BC, 17150BF, 17159BD, 17178AC, 17269BD, and 17312BB.
Class II
The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2 calibrators. The negative bias observed for patient, QC, and calibrator samples, when using any of the listed lots, has the potential to impact the interpretation of gentamicin results during therapeutic monitoring at concentrations up to approximately 2.5 ¿g/mL [5.4 ¿mol/L].
Lots 17135BC 17150BF 17159BD 17178AC 17269BD 17312BB
Nationwide distribution. Foreign distribution to Argentina, Australia, Brazil, Bahamas, Canada, Germany, India, Japan, Myanmar, New Zealand, Saudi Arabia, Taiwan, and South Africa.
Voluntary: Firm initiated