Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2019 - (Recall #: Z-1913-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

ADVIA Chemistry Creatine Kinase (CK_L) reagent, Product Code 10729780, UDI Number: 00630414008943 - Product Usage: The ADVIA¿ Chemistry Creatine Kinase (CK_L) assay is for in vitro diagnostic use in the quantitative determination of creatine kinase activity in human plasma (lithium heparin) or serum on ADVIA Chemistry XPT systems. The assay can be used to aid in the diagnosis and treatment of myocardial infarction and muscle diseases, such as Duchenne progressive muscular dystrophy.

Product Classification:

Class II

Date Initiated: May 3, 2019
Date Posted: July 10, 2019
Recall Number: Z-1913-2019
Event ID: 82980
Reason for Recall:

A steady upward trend in blank (u) absorbance was observed

Status: Terminated
Product Quantity: 3,386
Code Information:

Serial Numbers: 465336,465663,468449

Distribution Pattern:

Worldwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated