Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2020 - (Recall #: Z-0052-2021)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in human plasma Catalog Number: RF627, SMN# 10471068
Class II
Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin compared with a control lot was +25%. A maximum bias of +34% in patient samples around the 99th percentile observed
Lot # EB0255 UDI: 00842768037726EB025520091110471068840
US Nationwide distribution.
Voluntary: Firm initiated