Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2020 - (Recall #: Z-2282-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

ADVIA Chemistry Enzymatic Creatinine_2, Siemens Material Number 10335869, UDI Number 00630414526690 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.

Product Classification:

Class II

Date Initiated: May 4, 2020
Date Posted: June 17, 2020
Recall Number: Z-2282-2020
Event ID: 85602
Reason for Recall:

The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.

Status: Terminated
Product Quantity: 1006
Code Information:

Lot #s 53149, 53150, 53151, 53152, 53941, 53942, 53943, 53944, 54354, 54371, 54372, 54373, 54374

Distribution Pattern:

Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.

Voluntary or Mandated:

Voluntary: Firm initiated