Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2020 - (Recall #: Z-2747-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US
Product Classification:
Class II
Date Initiated: June 30, 2020
Date Posted: August 12, 2020
Recall Number: Z-2747-2020
Event ID: 85980
Reason for Recall:
If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.
Status: Terminated
Product Quantity: 1800 units
Code Information:
DCA Vantage Handheld Barcode Scanner #11306432 Model: DS4308 - HC0015BZZWW
Distribution Pattern:
Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated