Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0181-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Advia Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11319151
Product Classification:
Class II
Date Initiated: September 20, 2021
Date Posted: November 3, 2021
Recall Number: Z-0181-2022
Event ID: 88784
Reason for Recall:
The firm has confirmed the potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c results may affect consideration of intervention.
Status: Ongoing
Product Quantity: 246 (US); 689 (OUS)
Code Information:
All lots impacted
Distribution Pattern:
Worldwide distribution - US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated