Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2021 - (Recall #: Z-2028-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Atellica IM 1600 Analyzer- multi-component system for in vitro diagnostic testing of clinical specimens SMN 11066000
Class II
Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject to high pressure or high wear and may impact the result accuracy of any assay.
Effected serial range: IH00876 IH01451 and IRH00258xxxx IRH01169xxxx Note: xxxx in serial number is any number. UDI: 00630414002026
Nationwide Foreign: Country Name Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile Colombia Croatia Cura¿ao Cyprus Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Greece Hungary India Indonesia Iraq Ireland Israel Italy Japan Kenya Kuwait Latvia Libya Lithuania Malaysia Mexico Morocco Netherlands New Zealand Norway Oman P.R. China Pakistan Peru Philippines Poland Portugal Qatar Republic Korea Romania Russian Fed. Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. Uganda Ukraine United Kingdom Uruguay Vatikancity Vietnam
Voluntary: Firm initiated