Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0449-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Atellica CH 930 Analyzer with Software version: V1.25.1 and lower- An automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11067000

Product Classification:

Class II

Date Initiated: November 18, 2021
Date Posted: January 12, 2022
Recall Number: Z-0449-2022
Event ID: 89160
Reason for Recall:

(1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay Correlation factor; (2) On Board Stability (OBS) Not Updating with Manual Change for Open Channel Assays may continue to use the reagent past its expiration date leading to potential erroneous patient sample results

Status: Ongoing
Product Quantity: 2,980 units
Code Information:

Software version: V1.25.1 and lower. UDI: 00630414002163

Distribution Pattern:

Nationwide Foreign: AU DE AE TR BG CL IN CZ BR ES KE CN PL SA BE IQ IT CW SE AR CO GR SK EE KR JP UY DK FR GB VN CA BD MX TH DZ PT RU ZA PK HU NL VA NO AT IR LV EG PH SG MY IL FI HR MA KW CH TW BH AL BF UG NZ OM HK QA PE EC IE GP LY LT

Voluntary or Mandated:

Voluntary: Firm initiated