Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2022 - (Recall #: Z-1015-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number 10381194, catalog number L2KCG6]
Class II
Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.
8 production lots: 463, 464, 466, 467, 469, 470, 471, 472, [UDI for 200 tests: (01)00630414961132(10)463(17)20220228, (01)00630414961132(10)464(17)20220331, (01)00630414961132(10)466(17)20220331, (01)00630414961132(10)467(17)20220531, (01)00630414961132(10)469(17)20220731, (01)00630414961132(10)470(17)20220930, (01)00630414961132(10)471(17)20221130, (01)00630414961132(10)472(17)20230228] [UDI for 600 tests: (01)00630414961149(10)463(17)20220228, 01)00630414961149(10)464(17)20220331, (01)00630414961149(10)466(17)20220331, (01)00630414961149(10)467(17)20220531, (01)00630414961149(10)469(17)20220731, (01)00630414961149(10)470(17)20220930, (01)00630414961149(10)471(17)20221130, (01)00630414961149(10)472(17)20230228]
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, ID, IN, MN, NC, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TX, VA, WI, WY and OUS: 81 countries.
Voluntary: Firm initiated