Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2022 - (Recall #: Z-1816-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
Dimension Magnesium (MG) Flex reagent cartridge Siemens Material Number (SMN)/REF (Catalog Number):10444963/DF57
Class II
Dimension Magnesium (MG) Flex reagent cartridge lots may exhibit imprecision for Quality Control (QC) and produce Abnormal Reaction flags. The issue is not always detected by QC and erroneous results may be produced in the absence of an Abnormal Reaction flag. Based on customer data, imprecision leading to erroneous, unflagged patient results with a bias of -15 to -59% may occur
Lot Number: UDI MG lot FA2350 10444963 00842768014185FA235022121610444963840; MG lot FA2356 10444963 00842768014185FA235622122210444963840; MG lot GA2363 10444963 00842768014185GA236322122910444963840; MG lot FA3019 10444963 00842768014185FA301923011910444963840
Nationwided Foreign: Afghanistan¿¿¿¿ Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Bulgaria¿¿¿¿¿¿¿ Canada Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Egypt¿¿¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hungary¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Jordan¿¿¿¿¿¿¿¿¿ Kazakhstan¿¿¿¿¿ Kenya¿¿¿¿¿¿¿¿¿¿ Kuwait¿¿¿¿¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Lebanon¿¿¿¿¿¿¿¿ Macedonia¿¿¿¿¿¿ Mali¿¿¿¿¿¿¿¿¿¿¿ Malta¿¿¿¿¿¿¿¿¿¿ Montenegro¿¿¿¿¿ Morocco¿¿¿¿¿¿¿¿ Nepal¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Qatar¿¿¿¿¿¿¿¿¿¿ Romania¿¿¿¿¿¿¿¿ Russian¿Fed.¿¿¿ Saudi¿Arabia¿¿¿ Senegal¿¿¿¿¿¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Slovenia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Sudan¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ U.A.E.¿¿¿¿¿¿¿¿¿ Uganda¿¿¿¿¿¿¿¿¿ Ukraine¿¿¿¿¿¿¿¿ United¿Kingdom¿ Uzbekistan¿¿¿¿¿
Voluntary: Firm initiated