Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2023 - (Recall #: Z-1270-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Atellica CH Triglycerides (concentrated)-In vitro diagnostic use in the quantitative measurement of triglycerides in human serum and plasma (lithium heparin, potassium EDTA) using the Atellica CH Analyzer SMN: 11097591

Product Classification:

Class II

Date Initiated: January 27, 2023
Date Posted: September 6, 2023
Recall Number: Z-1270-2023
Event ID: 91665
Reason for Recall:

Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results

Status: Ongoing
Product Quantity: 11824 units
Code Information:

UDI: 00630414596495 All Lots

Distribution Pattern:

Nationwide Foreign: Country Albania Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Croatia Cura¿ao Czech Republic Denmark Ecuador Egypt Finland France Germany Greece Guadeloupe Hong Kong Hungary India Ireland Israel Italy Japan Kuwait Libya Malaysia Mexico Netherlands New Zealand Norway Oman Pakistan Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vatikancity Vietnam

Voluntary or Mandated:

Voluntary: Firm initiated