Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2023 - (Recall #: Z-2474-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Atellica¿ CH HDL Cholesterol- In vitro diagnostic use in the quantitative determination of HDL cholesterol in human serum and plasma (lithium heparin, EDTA, sodium Heparin) Test Code: HDLC Siemens Material Number (SMN): 11537213

Product Classification:

Class II

Date Initiated: July 10, 2023
Date Posted: September 6, 2023
Recall Number: Z-2474-2023
Event ID: 92800
Reason for Recall:

Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿ CH Total Bilirubin_2 (TBil_2), Atellica¿ CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica CH HDL Cholesterol (HDLC)

Status: Ongoing
Product Quantity: 18524 units
Code Information:

Unique Device Identification (UDI): 00630414610832 All lot numbers

Distribution Pattern:

US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam

Voluntary or Mandated:

Voluntary: Firm initiated