Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2024 - (Recall #: Z-1304-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
DCA Systems HbA1c Reagent Kit
Class II
The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.
Siemens Material Number: 10311134; UDI/DI: 00630414532806; Lot#: 9640093, 0645103, 0648103, 0649103, 0654103 . Expanded Lots (02/28/2024): 0637093, 0641093, 0642093, 0647103, 0653103, 0655103;
Worldwide distribution - US Nationwide and the country of Taiwan.
Voluntary: Firm initiated