Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2024 - (Recall #: Z-3169-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Atellica CH Microalbumin_2 (¿ALB_2)-For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers

Product Classification:

Class II

Date Initiated: August 15, 2024
Date Posted: September 25, 2024
Recall Number: Z-3169-2024
Event ID: 95218
Reason for Recall:

Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.

Status: Ongoing
Product Quantity: 18483 units
Code Information:

UDI-DI: 00630414596310 Lot Numbers: 232033, 232128, 232137, 232146, 232147, 242149, 242150, 242365

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Canada, Chile, Colombia, Croatia, Cura¿ao, St. Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong¿Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lithuania, Malaysia, Maldives, Mexico, Morocco, Netherlands, New¿Zealand, Norway, Oman, P.R.¿China, Pakistan, Paraguay, Peru, Poland, Portugal, Qatar, Republic¿Korea, Romania, Russian¿Fed., Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South¿Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, U.A.E, United¿Kingdom, Uruguay, Vietnam, Zaire. ¿¿¿¿¿¿

Voluntary or Mandated:

Voluntary: Firm initiated