Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2025 - (Recall #: Z-2027-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;
Product Classification:
Class II
Date Initiated: May 22, 2025
Date Posted: July 2, 2025
Recall Number: Z-2027-2025
Event ID: 96952
Reason for Recall:
Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
Status: Ongoing
Product Quantity: 1,625 units (1450 US, 175 OUS)
Code Information:
Siemens Material Number (SMN): 11099300; UDI-DI: 00630414595771; Lot Number: 0000195412;
Distribution Pattern:
Worldwide distribution - US Nationwide and the countries of Canada, Bahamas.
Voluntary or Mandated:
Voluntary: Firm initiated