Siemens Healthcare Diagnostics: Medical Device Recall in 2014 - (Recall #: Z-2643-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

ADVIA Chemistry A1c_3 CAL (Calibrator) for use on the ADVIA Chemistry Systems, REF/Catalog Number/Siemens Material Number (SMN) 10491408, IVD ADVIA Chemistry A1c_3 Calibrator Information: For in vitro diagnostic use in the calibration of A1c_3 (A1c_3 and A1c_3M) and Total Hemoglobin_3 (tHb_3 and tHb_3M) methods on ADVIA Chemistry systems. The ADVIA Chemistry systems are automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine in random access and batch modes for in vitro diagnostic use.

Product Classification:

Class II

Date Initiated: August 6, 2014
Date Posted: September 24, 2014
Recall Number: Z-2643-2014
Event ID: 68871
Reason for Recall:

The ADVIA Chemistry Systems A1c_3 method may exhibit a percent bias of -9% to +11% for hemoglobin (Hb)A1c samples when using A1c_3 Calibrator lots 3HD044 and 3LD068. Depending on quality control limits, this issue may not have been detected. NOTE: The percent bias range is a percentage of the HbA1c result. It is NOT absolute HbA1c units.

Status: Terminated
Product Quantity: Domestic: 1122 kits; Foreign: 236 kits
Code Information:

Lots 3HD044 (exp. 11/01/2015), and 3LD068 (exp. 01/01/2016)

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Austria, Brazil, Canada, China, Colombia, Croatia, Cyprus, Denmark, Estonia, Finland, France, Germany, Greece, Guatemala, India, Indonesia, Istanbul, Italy, Lithuania, Mexico, Paraguay, Peru, Poland, Romania, Russian Federation, Serbia, Singapore, Slovakia, South Korea, Spain, Switzerland, Thailand, Uruguay, Vietnam, and Zambia.

Voluntary or Mandated:

Voluntary: Firm initiated