Siemens Medical Solutions USA, Inc.: Medical Device Recall in 2016 - (Recall #: Z-0667-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Z6Ms transesophageal transducers used on ACUSON SC2000 ultrasound systems, Z6Ms model 10436113. Radiology: The ACUSON SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Abdominal Intraoperative, Intraoperative Neurological, Neurological-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctive with other medical data obtained by a physician for clinical diagnosis purposes.

Product Classification:

Class II

Date Initiated: November 7, 2016
Date Posted: December 7, 2016
Recall Number: Z-0667-2017
Event ID: 75639
Reason for Recall:

Firm received reports about "acquisition errors" and the deterioration of material covering the articulating section of the Z6Ms transesophageal transducer.

Status: Terminated
Product Quantity: 33
Code Information:

Model number 10436113 -Z6Ms Transesophageal tranducer. US Serial numbers: 63229007 63229008 63432010 63530518 63322007 63230009 63260584 63530516 63432007 63319012 63332004 63325007 63328003 63328007 63332002 63029015 63329017 63230003. Foreign serial numbers: 63260585 63230002 63029009 63329018 63229011 63432015 63229012 63430524 63029016 63360611 63360612 63432017.

Distribution Pattern:

Worldwide Distribution - US (nationwide) Internationally to AUSTRIA, GERMANY, SOUTH AFRICA, NETHERLANDS, MAURITIUS, MEXICO, and ITALY, JAPAN, SWEDEN

Voluntary or Mandated:

Voluntary: Firm initiated