Smith & Nephew, Inc.: Medical Device Recall in 2018 - (Recall #: Z-1921-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.
Product Classification:
Class II
Date Initiated: May 2, 2018
Date Posted: May 30, 2018
Recall Number: Z-1921-2018
Event ID: 80050
Reason for Recall:
Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.
Status: Terminated
Product Quantity: 13 units
Code Information:
Lot Numbers: 17LM22019, 17LM10934, 17LM05900
Distribution Pattern:
US nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated