Smiths Medical Asd Inc: Medical Device Recall in 2024 - (Recall #: Z-0740-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

1050 STOPCOCK LEFT ROTATOR OFF HANDLE, List Number MX1431MRL

Product Classification:

Class II

Date Initiated: November 10, 2023
Date Posted: January 24, 2024
Recall Number: Z-0740-2024
Event ID: 93506
Reason for Recall:

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

Status: Ongoing
Product Quantity: 31,685 total
Code Information:

Lot Number: 4387166

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Canada, UAE.

Voluntary or Mandated:

Voluntary: Firm initiated