Smiths Medical Asd Inc: Medical Device Recall in 2024 - (Recall #: Z-0745-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
20IN (50.8CM) BRAIDED INJ LINE, ROTATING ADAPT, List Number MX682BR
Product Classification:
Class II
Date Initiated: November 10, 2023
Date Posted: January 24, 2024
Recall Number: Z-0745-2024
Event ID: 93506
Reason for Recall:
There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
Status: Ongoing
Product Quantity: 31,685 total
Code Information:
Lot Number: 4398428 4404690
Distribution Pattern:
Worldwide distribution - US Nationwide and the countries of Canada, UAE.
Voluntary or Mandated:
Voluntary: Firm initiated