Sorin Group Italia SRL - CRF: Medical Device Recall in 2017 - (Recall #: Z-3230-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Platinium CRT DF1 1711, Model Number TDF033U, UDI/GTIN 08031527015538, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).

Product Classification:

Class II

Date Initiated: July 24, 2017
Date Posted: October 4, 2017
Recall Number: Z-3230-2017
Event ID: 77983
Reason for Recall:

There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.

Status: Terminated
Product Quantity: 154 units
Code Information:

All Serial Numbers

Distribution Pattern:

US Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated