Sorin Group USA, Inc.: Medical Device Recall in 2012 - (Recall #: Z-0158-2013)
See the recall detail below. You can also see other recalls from the same firm in 2012.
Gish 3/8 x 1/2" Reducer Connector with Luer fitting, Item Numbers: EC2180S, Lot Numbers: 1108900099, 1119900091, 1203700129, 1208600013, 1219200107, Exp. Dates: 3/31/14, 7/31/14, 2/28/15, 3/31/15, 7/31/15. Product Usage: The connectors are indicated for use in connecting flexible tubing used in extracorporeal blood or fluid circuits. The connectors are distributed as sterile stand alones and bulk non-sterile.
Class II
Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.
Item Numbers: EC2180S, Lot Numbers: 1108900099, 1119900091, 1203700129, 1208600013, 1219200107, Exp. Dates: 3/31/14, 7/31/14, 2/28/15, 3/31/15, 7/31/15.
US Nationwide Distribution in the states of: AL, AR, CA, CO, CT, FL, IL, MI, MN, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA and WI.
Voluntary: Firm initiated