Sorin Group USA, Inc.: Medical Device Recall in 2012 - (Recall #: Z-0166-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

SCP Pump Control Panel, Product Code: 60-02-15, Serial Numbers: 60S09121, 60S09123-60S09129, 60S09136-60S09139. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuit for typical durations of six hours or less.

Product Classification:

Class II

Date Initiated: September 20, 2012
Date Posted: November 7, 2012
Recall Number: Z-0166-2013
Event ID: 63381
Reason for Recall:

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.

Status: Terminated
Product Quantity: 12 units (3 units USA and 9 units OUS)
Code Information:

Product Code: 60-02-15, Serial Numbers: 60S09121, 60S09123-60S09129, 60S09136-60S09139.

Distribution Pattern:

Worldwide distribution: USA (nationwide) including states of : MI and NY; and countries of: Belgium, Canada, Japan, India, Russia, United Arab Emirates.

Voluntary or Mandated:

Voluntary: Firm initiated