Sorin Group USA, Inc.: Medical Device Recall in 2015 - (Recall #: Z-2081-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

Sorin Stockert Heater-Cooler 3T, 200 V / 50Hz / 60 Hz Temperature control for extracorporeal perfusion of durations up to 6 hours.

Product Classification:

Class II

Date Initiated: June 15, 2015
Date Posted: July 22, 2015
Recall Number: Z-2081-2015
Event ID: 71593
Reason for Recall:

Potential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.

Status: Terminated
Product Quantity: 131
Code Information:

Product code 16-02-95 Serial number 16S12004-16S15385

Distribution Pattern:

Worldwide Distribution - US (nationwide) and Internationally to AL, AT, BE, BG, CH, CY, CZ, DE, DK, EE, ES, FI, FR, GB, GR, HR, HU, IE, IT, LT, LU, LV, MC, MK, MT, NL, NO, PL, PT, RO, RS, SE, SK, AE, AR, AU, AZ, BD, BH, BR, BY, CA, CL, CN, CO, CR, DZ, EC, EG, ET, GE, GY, HK, ID, IL, IN, IQ, IR, JO, JP, KR, KW, KZ, LB, LK, LY, MA, MN, MU, MX, MY, NG, NP, NZ, OM, PA, PE, PH, PK, PR, PS, QA, RE, RU, SA, SG, SV, SY, TH, TN, TR, TT, TW, UA, VN, ZA.

Voluntary or Mandated:

Voluntary: Firm initiated