Spacelabs Healthcare, Inc.: Medical Device Recall in 2023 - (Recall #: Z-1503-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
Ultraview SL (UVSL) Command Module, Model 91496
Class II
Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.
Software version 2.03.06 and earlier and software version 2.04 and later.
US Nationwide distribution in the states of SD, KY, MO, WA, OH, NC, CA, OR, TX, GA, IL, OK, SC, PA, AL, KS, FL, ID, MI, MN, IA, NJ, MD, VA, AR, CT, NE, LA, AK, MS, ME, WV, NY, AZ, TN, WY, IN, MA, MT, CO, DE, ND, WI, NV, RI, NH, UT.
Voluntary: Firm initiated