Spacelabs Healthcare, Inc.: Medical Device Recall in 2025 - (Recall #: Z-0169-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0893-01 Software Version: N/A Product Description: 16 inch DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with the 91390 and 91393

Product Classification:

Class II

Date Initiated: September 5, 2025
Date Posted: October 22, 2025
Recall Number: Z-0169-2026
Event ID: 97616
Reason for Recall:

Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

Status: Ongoing
Product Quantity: 16898
Code Information:

Lot Code: UDI-DI for products that are used with the display cables. (01)10841522131479 (Xprezzon), (01)10841522131455 (Qube)

Distribution Pattern:

Worldwide - United States Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of AFG, ARE, ARG, AUS, AUT, BGD, BHR, BOL, BRA, BRN, CAN, CHE, CHL, CHN, COL, CRI, CYP, CZE, DOM, DZA, ECU, EGY, ESP, FRA, GBR, GRC, GTM, GUM, HKG, IDN, IND, IRL, IRN, ISR, ITA, JOR, KWT, LBN, MAR, MEX, MYS, NLD, NZL, OMN, PAK, PAN, PER, PHL, PLS, POL, PRI, PRT, PRY, QAT, ROM, SAU, SGP, SLV, SRB, THA, TTO, TUN, TUR, TWN, VNM.

Voluntary or Mandated:

Voluntary: Firm initiated