Spacelabs Healthcare, Ltd.: Medical Device Recall in 2020 - (Recall #: Z-0563-2021)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Sentinel V10.x & V11.x, Models: 98200 & 98201 - Product Usage: designed using the latest programming technology and is easy to install and to use.
Class II
During system upgrade incorrect comments may be added to existing tests. In such instance: Resting ECG and ABP tests which did not contain comments at the time of the upgrade, may have unrelated comments added. Existing comments will not be overwritten or removed. Rhythm ECG Tests may have comments overwritten with unrelated comments. Any tests created after the upgrade are not affected.
Systems that have been upgraded from Version 9 or below to Version 10 or above.
Worldwide distribution - U.S. Nationwide distribution including in the states of PR, MN, NE, FL, IL, CA, NY, ND, AR, WI, SD, TN, MO, MA, TX, MT, OH, MI, IA, NJ, CT, NC, UT, WV, GA, CO, PA, AL, DC, NV, DE, WA, OR, IN, SC, HI, MS, KS, ID, MD, LA, NH, VA, ME, AK, VT, OK, AZ, KY, NM, RI and the countries of AFGHANISTAN, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BELGIUM, BOLIVIA, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALY, KENYA, KOREA, REPUBLIC OF, KUWAIT, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PALESTINE, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, Republic of Macedonia, SAUDI ARABIA, Serbia, SINGAPORE, Slovakia, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VENEZUELA.
Voluntary: Firm initiated