Specialty Appliance Works, Inc.: Medical Device Recall in 2015 - (Recall #: Z-1504-2015)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
TS Herbst Nuts with Backing (10/pk) and TS Herbst Nuts Universal (10/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.
Product Classification:
Class II
Date Initiated: March 9, 2015
Date Posted: April 29, 2015
Recall Number: Z-1504-2015
Event ID: 70787
Reason for Recall:
The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.
Status: Terminated
Product Quantity: 14,428 Universal Nuts
Code Information:
Product No. 90869 and 90875
Distribution Pattern:
Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.
Voluntary or Mandated:
Voluntary: Firm initiated