Spectranetics Corporation: Medical Device Recall in 2022 - (Recall #: Z-0221-2023)
See the recall detail below. You can also see other recalls from the same firm in 2022.
Philips Laser System, REF: LAS-100. Used in minimally invasive interventional procedures within the cardiovascular system, and for the removal of problematic pacemaker and defibrillator cardiac leads.
Class II
Laser System may detect Error 106 or Error 108 System Failures that will cause the system to enter non-recoverable safe state. The user will not be able to proceed until the error is cleared. May result in initiation of treatment delay and/or treatment beyond initial scope and/or unable to treat patient.
UDI-00813132027452, Software Version: 0.5.0.3, Serial Numbers: 100000, 100001, 100002, 100003, 100004, 100005, 100006, 100007, 100008, 100009, 100010, 100011, 100012, 100013, 100014, 100015, 100016, 100017, 100018, 100019, 100020, 100021, 100022, 100023, 100024, 100025, 100026, 100027, 100028, 100029, 100030, 100033, 100034, 100035, 100036, 100037, 100038, 100039, 100040, 100041, 100042, 100043, 100044, 100045, 100046, 100047, 100048, 100049, 100050, 100051, 100052, 100053, 100054, 100055, 100058, 100059, 100060, 100061, 100062, 100063, 100064, 100065, 100066, 100067, 100068, 100069, 100070, 100071, 100072, 100073, 100074, 100075, 100076, 100077, 100078, 100079, 100080, 100081, 100082, 100083, 100084, 100085, 100086, 100087, 100088, 100089, 100090, 100091, 100092, 100093, 100094, 100095, 100096, 100097, 100098, 100099, 100100, 100101, 100102, 100103, 100104, 100105, 100106, 100107, 100108, 100112, 100113, 100114, 100115, 100116, 100117, 100118, 100120
US: LA, MO, TN, CA, NC, PA, HI, AZ, CO, OK, FL, TX, MN, MI, VA, CT, NY, MD, IL, NJ, MA, KS, ME, AR, IA, IN, GA, ID, OR. OUS: DE, GB, ES, SE, IT, RS, GR, IN, FR, IL, DK, SA
Voluntary: Firm initiated