SPINEART SA: Medical Device Recall in 2023 - (Recall #: Z-1426-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, JLT-PL 02 16-S; JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE LARGE, REF: JLT-PL 04 08-S, JLT-PL 04 10-S, JLT-PL 04 12-S, JLT-PL 04 14-S, JLT-PL 04 16-S, JLT-PL 04 18-S

Product Classification:

Class II

Date Initiated: March 16, 2023
Date Posted: May 3, 2023
Recall Number: Z-1426-2023
Event ID: 91973
Reason for Recall:

JULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could cause surgical delay.

Status: Ongoing
Product Quantity: 544
Code Information:

REF/UDI-DI/Lot: JLT-PL 02 08-S/07640178987388/4-8826, 5-4142, 5-5377, 5-0643, 5-2431; JLT-PL 02 10-S/07640178987395/6-5488, 6-6556, 5-0644, 6-4734, 6-4735, 5-5378, 6-3211; JLT-PL 02 12-S/07640178987401/6-5489, 6-6558, 6-7645, 6-8734, 6-4732, 4-8828, 5-5379, 6-4733; JLT-PL 02 14-S/07640178987418/5-8439, 4-8829, 6-0714, 5-5380; JLT-PL 02 16-S/07640178987425/4-8830, 5-4145, 6-3217, 6-0726, 6-5490; JLT-PL 04 08-S/07640178987456/4-8845, 5-4146, 6-0725; JLT-PL 04 10-S/07640178987463/4-8831, 5-4147, 6-0722; JLT-PL 04 12-S/07640178987470/4-8832, 5-4148, 5-8438, 6-0721, 6-3209; JLT-PL 04 14-S/07640178987487/4-3861, 4-8833, 5-4149; JLT-PL 04 16-S/07640178987494/5-4150; JLT-PL 04 18-S/07640178987500/5-4151

Distribution Pattern:

US Nationwide distribution in the states of NH, TX, CA, VA, IL, RI, NC, AZ.

Voluntary or Mandated:

Voluntary: Firm initiated