SPS Sterilization, Inc: Medical Device Recall in 2018 - (Recall #: Z-3041-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Vascular Solutions Venture Rx Catheter, 6F, resterilized.

Product Classification:

Class II

Date Initiated: March 7, 2018
Date Posted: September 19, 2018
Recall Number: Z-3041-2018
Event ID: 80317
Reason for Recall:

The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.

Status: Terminated
Product Quantity: 3 units total
Code Information:

Code Number 5820-R; Lot Number 570183 (Serial Number 99804132016-07); Lot Number 583022 (Serial Number 17304202017-07, 17304202017-08)

Distribution Pattern:

US distribution to AZ, CA, FL, OH, PA, TX, VA, WV

Voluntary or Mandated:

Voluntary: Firm initiated