St. Jude Medical, Cardiac Rhythm Management Division: Medical Device Recall in 2022 - (Recall #: Z-0988-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2022.
Data Source: FDA.
Product Description:
Gallant DR Implantable Cardioverter Defibrillator REF # CDDRA500Q
Product Classification:
Class II
Date Initiated: March 10, 2022
Date Posted: April 27, 2022
Recall Number: Z-0988-2022
Event ID: 89825
Reason for Recall:
Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.
Status: Ongoing
Product Quantity: 1
Code Information:
REF# CDDRA500Q SN# 111018237, Exp Date: 31-Oct-2022 Material# 600115496 UDI# 05415067032003
Distribution Pattern:
Worldwide distribution - US Nationwide distribution in the states of FL, KY and the country of France, Czech Republic.
Voluntary or Mandated:
Voluntary: Firm initiated