St. Jude Medical, Cardiac Rhythm Management Division: Medical Device Recall in 2022 - (Recall #: Z-0989-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
Gallant VR Implantable Cardioverter Defibrillator REF # CDVRA500Q
Class II
Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.
REF# CDVRA500Q SN: 111009251, Exp Date : 31-Jul-2022, Material # 100173801; SN # 810000544, Exp Date: 31-Oct-2022, Material # 600115400 UDI# 05415067031990
Worldwide distribution - US Nationwide distribution in the states of FL, KY and the country of France, Czech Republic.
Voluntary: Firm initiated