St Jude Medical Inc.: Medical Device Recall in 2019 - (Recall #: Z-2073-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
ELLIPSE DR, REF: CD2377-36QC Implantable cardioverter defibrillators (ICDs)
Class I
Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.
Serial Number: 1282011, 1282019, 1282065, 1282069, 1282424, 1282458, 1282469, 1282474, 1282596, 1282602, 1282609, 1282622, 1282625, 1282638, 1282641, 1282648, 9829389, 9831310, 9831509, 9831515, 9831516
Worldwide distribution. US nationwide, and countries: DE, FR, IT, ES, NL, PT, SE, HU, DK, PL, IL, GP, DZ, CH, SA, SK, LU, NZ, VN, and NO
Voluntary: Firm initiated