St Jude Medical: Medical Device Recall in 2014 - (Recall #: Z-0177-2015)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
St. Jude Medical, Fixed Curve Response 6F Electrophysiology Catheter, CSL, Model 401353
Product Classification:
Class II
Date Initiated: October 14, 2014
Date Posted: November 12, 2014
Recall Number: Z-0177-2015
Event ID: 69512
Reason for Recall:
St. Jude Medical is performing a voluntary recall of batch 4671318, model number 401353 of the Response Fixed Curve DECA 6F 65cm CSL 2/8 Catheter. Specifically, the firm has identified that every device packaged within batch 4671318 contains the incorrect catheter (Supreme FC QUAD 5F catheter) within the packaging. To date, no serious injuries have occurred as a result of this issue
Status: Terminated
Product Quantity: 26
Code Information:
Batch # 4671318
Distribution Pattern:
US Distribution including the states of CA, GA, NJ, NY, OR and PA.
Voluntary or Mandated:
Voluntary: Firm initiated