Staar Surgical AG: Medical Device Recall in 2026 - (Recall #: Z-1480-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2026.

Data Source: FDA.
Product Description:

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Product Classification:

Class II

Date Initiated: January 21, 2026
Date Posted: March 11, 2026
Recall Number: Z-1480-2026
Event ID: 98397
Reason for Recall:

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.

Status: Ongoing
Product Quantity: 7
Code Information:

UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988

Distribution Pattern:

International distribution to the countries of India, Iran, Korea, Saudi Arabia.

Voluntary or Mandated:

Voluntary: Firm initiated