Staar Surgical AG: Medical Device Recall in 2026 - (Recall #: Z-1480-2026)
See the recall detail below. You can also see other recalls from the same firm in 2026.
EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
Class II
Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988
International distribution to the countries of India, Iran, Korea, Saudi Arabia.
Voluntary: Firm initiated