Product Description:
Alaris Pump Module, Model 8100, serviced/repaired by Step-Har Medical with an impacted bezel repair part
Product Classification:
Class I
Date Initiated: February 26, 2021
Date Posted: May 19, 2021
Recall Number: Z-1543-2021
Event ID: 87773
Reason for Recall:
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Status: Terminated
Product Quantity: 2
Code Information:
Serial Numbers: 12747983 and 9919591
Distribution Pattern:
IA
Voluntary or Mandated:
Voluntary: Firm initiated