Steris Corporation: Medical Device Recall in 2013 - (Recall #: Z-0141-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Harmony Equipment Management System, Model #603X/80XX, STERIS Corporation, Mentor, OH Harmony Equipment Management Systems (EMS) are used in hospital facilities to provide ready access to medical gasses, electrical power and audio/visual data services, in addition to managing surgical support equipment such as endoscopic equipment, cautery devices and insufflators.

Product Classification:

Class II

Date Initiated: August 15, 2013
Date Posted: November 13, 2013
Recall Number: Z-0141-2014
Event ID: 66138
Reason for Recall:

Dr¿ger Medical GmbH has notified STERIS of the potential for a break in the drive screw contained in the motorized ceiling supply unit, a component of the Harmony EMS system. The ceiling supply unit controls the motorized adjustment of the height and rotation of the supply column. Should a break occur in the drive screw, there is potential for the adjustable arm to drop down, causing the EMS sys

Status: Terminated
Product Quantity: 127 units
Code Information:

Serial numbers 0422503114 - 0426604096

Distribution Pattern:

Nationwide distribution: USA including the states of Alabama, Arkansas, Florida, Georgia, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Nebraska, New Jersey, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Texas, Virginia, Washington and Wisconsin.

Voluntary or Mandated:

Voluntary: Firm initiated