Steris Corporation: Medical Device Recall in 2021 - (Recall #: Z-0790-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Vis-U-All High Temp 4.5"x11" Heat Seal Pouch 200 pouches per box; 5 boxes per case Model Number: 883411
Product Classification:
Class II
Date Initiated: December 2, 2020
Date Posted: January 13, 2021
Recall Number: Z-0790-2021
Event ID: 86942
Reason for Recall:
Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Status: Terminated
Product Quantity: 53 cases
Code Information:
Lot Number: 200421A, 200521A
Distribution Pattern:
Nationwide Foreign:
Voluntary or Mandated:
Voluntary: Firm initiated