Steris Corporation: Medical Device Recall in 2021 - (Recall #: Z-1149-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

OT1000 Series Orthopedic Surgical Tables

Product Classification:

Class II

Date Initiated: January 21, 2021
Date Posted: March 3, 2021
Recall Number: Z-1149-2021
Event ID: 87230
Reason for Recall:

On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.

Status: Terminated
Product Quantity: 272 tables total (Domestic: 227 tables, Foreign 45 tables)
Code Information:

Model No.: OT1000, OT1100, OT1200

Distribution Pattern:

Domestic: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL IN, IW, KS, KY, LA, ME, MD, MI, MN, MO, NE, NV, NJ, NY, NC, ND, OH, OK, OR, PA, PR, SC, TX, UT, VT, VA, WA, WV, WI. Foreign: Canada, South Korea, Panama, Qatar, Thailand.

Voluntary or Mandated:

Voluntary: Firm initiated