Stryker Communications: Medical Device Recall in 2020 - (Recall #: Z-1602-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000

Product Classification:

Class II

Date Initiated: October 11, 2019
Date Posted: April 8, 2020
Recall Number: Z-1602-2020
Event ID: 85083
Reason for Recall:

The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.

Status: Terminated
Product Quantity: 1415 Surgical Lights
Code Information:

Serial Number Range: 091316 70016 through 083117-73252

Distribution Pattern:

US: Nationwide OUS: Canada

Voluntary or Mandated:

Voluntary: Firm initiated