Stryker Corporation: Medical Device Recall in 2023 - (Recall #: Z-2602-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Stryker ACM (Advanced Cement Mixing)System: ACM Kit w/ Femoral Breakaway Nozzle, Restrictors & Prox. Med. Press- Allows the user to mix bone cement with a constant high vacuum through the process. Catalog Number: 0306-705-000

Product Classification:

Class II

Date Initiated: August 7, 2023
Date Posted: September 27, 2023
Recall Number: Z-2602-2023
Event ID: 92841
Reason for Recall:

Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site

Status: Ongoing
Product Quantity: 231 units US; 21 units OUS
Code Information:

GTIN: 04546540055422 Lot Numbers: 22306012 22307012 22340012 22341012 22342012 23017012

Distribution Pattern:

Nationwide Foreign: Austria Belgium Canada France Germany Hong Kong Italy Malaysia Poland Singapore South Korea Spain Taiwan Thailand UK

Voluntary or Mandated:

Voluntary: Firm initiated