Stryker Corporation: Medical Device Recall in 2024 - (Recall #: Z-0367-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-001

Product Classification:

Class II

Date Initiated: October 4, 2024
Date Posted: November 20, 2024
Recall Number: Z-0367-2025
Event ID: 95502
Reason for Recall:

There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries

Status: Ongoing
Product Quantity: 2,620 eaches
Code Information:

GTIN: 07613327117219 Lot Number: 64117291

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.

Voluntary or Mandated:

Voluntary: Firm initiated